hit counter script

Advertisement

Quick Links

INSTRUCTION MANUAL
Air Mattress
cococia Fūka
7F11336300A2
Preface
Thank you for purchasing the Air Mattress cococia Fūka (hereinafter referred
to as the air mattress).
This instruction manual describes precautions and how to safely use and operate
the air mattress.
Before using this air mattress, please read this instruction manual to fully
understand the safe and proper operating procedures.
In addition to the person who uses this air mattress, carers (attendants,
nurses, caregivers, etc.) should also read this instruction manual.
After reading this instruction manual, store it in a place that can be easily
accessed for later reference.
Users and/or carers are requested to report information about any serious
incident related to this air mattress that occurs within the EU to the
manufacturer and the regulating authority of the member country of
residence.
'Serious incident' means any incident that directly or indirectly led, might have
led or might lead to any of the following:
a) the death of a user, carer or other person.
b) the temporary or permanent serious deterioration of a user's, carer's or
other person's state of health.
c) a serious public health threat.
Some of the descriptions in this instruction manual may differ from the product
you have purchased due to model refinements.
Values in the specifications such as operating ranges, dimensions, angles, and
weight are subject to normal manufacturing tolerances. With some exemptions,
words such as "approximately" and "about" are omitted for readability in this
instruction manual.
Contact us (see "Contact Information" on p. 32) for any queries.
PARAMOUNT BED CO., LTD.
cococia
Fūka

Advertisement

Table of Contents
loading

Summary of Contents for PARAMOUNT BED cococia Fuka

  • Page 1 Values in the specifications such as operating ranges, dimensions, angles, and weight are subject to normal manufacturing tolerances. With some exemptions, words such as “approximately” and “about” are omitted for readability in this instruction manual. Contact us (see “Contact Information” on p. 32) for any queries. PARAMOUNT BED CO., LTD.
  • Page 2: Table Of Contents

    CONTENTS ··························································· 1 INTENDED USE PART NAMES AND CHECKING PARTS ······················ 2 - 3 SAFETY PRECAUTIONS ·············································· 4 - 9 FUNCTIONS ································································ 10 11 - 14 INSTALLATION ··························································· ········································ 15 - 16 OPERATING PROCEDURES 17 - 20 MAINTENANCE AND INSPECTION ·······························...
  • Page 3: Part Names And Checking Parts

    PART NAMES AND CHECKING PARTS Keep the packing materials (cardboard boxes and vinyl bags) in which the air mattress and pump were delivered as they will be used for long-term storage and transportation. (See p. 23 and p. 24) <Air Mattress> Applied part * Part of ME EQUIPMENT that in normal use necessarily comes into...
  • Page 4 PART NAMES AND CHECKING PARTS <Pump> Handle Operation panel Power cable Power plug Blowing tube insertion port Cable clamp Case Power button Operation Panel Enlarged View Status display Back-raised mode button Max inflate mode button Hardness adjustment button Alternate pressure button Board hook KE-959E 2020-11...
  • Page 5: Safety Precautions

    SAFETY PRECAUTIONS To ensure safe and proper use, be sure to read this "SAFETY PRECAUTIONS" before use. The safety precautions described here are intended to ensure safe use of the product and prevent personal injury and property damage. Precautions are classified as either WARNING and CAUTION, which indicate the degree of hazard or injury that could result from improper use.
  • Page 6 Paramount Bed. Take care to prevent the air mattress from The use of products other than the specified protruding from the bed and slipping off it.
  • Page 7 WARNING When hooking the pump to the bed, hang it Do not allow dust to accumulate on the power over the footboard with the operation panel plug. facing the outside of the bed. Dust on the power plug can absorb moisture and conduct If it is hung anywhere else other than over the electricity, causing a short footboard or with the operation panel facing the...
  • Page 8 SAFETY PRECAUTIONS WARNING Do not allow the air mattress to be operated by Do not lie facedown. a person deemed incapable of understanding Lying facedown for a long time could lead to suffocation. mattress operation (a child aged 12 or under or a person with dementia).
  • Page 9 WARNING Disconnect the power plug from the outlet Inspect the air mattress and pump regularly when using electronic medical equipment. (see p. 20). Using the air mattress in conjunction with electronic or maintain basic safet and essential performance in regards to EMC the pump have an medical equipment (e.g.
  • Page 10 SAFETY PRECAUTIONS CAUTION Pay attention to the following items to make the Do not operate the air mattress while an object best use of excellent body pressure dispersion. is placed on it. Consult with an expert to prevent bedsores. The object may fall from the air mattress, possibly Bedsores cannot be prevented by this air mattress causing damage.
  • Page 11: Functions

    FUNCTIONS This air mattress has the following functions and setting modes, and, for example, the internal pressure of the air cells can be changed. For how to operate each function and how to switch modes, see p. 15. Hardness The internal pressure of the air mattress can be adjusted in five steps. adjustment function Alternate pressure Sets of three air cells alternately expand and contract repeatedly and release the...
  • Page 12: Installation

    INSTALLATION Use the following procedure to install the air mattress. 1. Installing the air mattress 2. Connecting the air mattress and pump 3. Installing pump 4. Laying power cable 1. Installing the air mattress This air mattress is a body pressure dispersion mattress that is used as an alternative to general-purpose mattresses.
  • Page 13 Install the bed on which the air mattress is placed at the required distance from the wall at the foot end. The required distance is the value calculated by adding 11 cm to the distance specified in the instruction manual for the bed.
  • Page 14 INSTALLATION 3. Installing pump Attach the board hook of the pump to the footboard of the bed and secure it. If the pump is used with a bed where the hook cannot be attached to the footboard, install the pump in a horizontal stable location at the foot end where it won't get in the way.
  • Page 15 4. Laying power cable Raise the bed to the highest position. Route the power cable on the side where the user does not get on or off the air mattress and the bed side table is not used. Power cable Route the power cable on the side where the user does not get on or off the air mattress and the over bed table is not used.
  • Page 16: Operating Procedures

    BEFORE USE 1. Operation panel part names of the pump and how to operate each function The setting of each function of the air mattress (see p. 10) can be changed by pressing an operation button on the operation panel. Operation buttons of the pump There are the following six buttons on the operation panel.
  • Page 17 2. How to lay the sheet When the air mattress has expanded, lay a sheet and mattress pad. Do not cover the CPR valve with a sheet, but store it between the headboard and the air mattress. Do not cover the CPR valve with a sheet. The CPR function may not be able WARNING to be used quickly in case of emergency.
  • Page 18: Maintenance And Inspection

    MAINTENANCE AND INSPECTION 1. Maintenance Always unplug the power plug before performing maintenance. Failure to WARNING do so may cause the air mattress to malfunction, possibly causing injury. If the pump is removed from the air mattress for maintenance, be careful that water or other liquid does not enter the coupling and blowing tube insertion port.
  • Page 19 [How to clean the top cover] Open the fasteners and remove the top cover from the bottom cover. Use soapy water (commercially available soap, such as neutral detergent diluted with water) to soak the cover to wash. Rinse with water until there is no residual detergent. Do not twist, but wring out gently without causing damage.
  • Page 20 MAINTENANCE AND INSPECTION How to maintain the pump [How to clean the pump] Periodically maintain the pump. Always unplug the power plug before maintenance and use the following procedure. Remove contaminants, if any (blood, feces, etc.). Use soapy water (commercially available soap, such as neutral detergent diluted with water) to wipe the pump to remove stains.
  • Page 21 2. Inspection Daily inspection Perform daily inspection for use while paying attention to the following points. Check Name Inspection items method Sinking extremely as shown in Figure 1. When the alternate pressure function is set to ON, some air cells are sinking. Figure 1 If the air mattress is sinking extremely, air Air mattress...
  • Page 22: Troubleshooting

    TROUBLESHOOTING Problems may not always indicate faults. Check the following once again before requesting repair. If the problem persists after checking and taking corrective action, or if the cause is unknown, stop using the air mattress immediately, disconnect the power plug from the outlet, and contact your distributor (see "Contact Information"...
  • Page 23 Symptoms and status Check Corrective action Since the max inflate mode is set to ON, the internal pressure of the air Is the max inflate mode lamp ON? mattress is high. When not necessary, set it to OFF. Air mattress is too hard Since the back-raised mode is set to ON, the internal pressure of the air Is the back-raised mode lamp ON?
  • Page 24: Long-Term Storage, Transportation, And Disposal

    LONG-TERM STORAGE, TRANSPORTATION, AND DISPOSAL When the air mattress will not be in use for an extended period of time, or when the air mattress will be packed and transported, pack the air mattress in a cardboard box and store or transport it according to the following "How to put away the air mattress."...
  • Page 25 (For 15-cm thickness) Put the pump in the cardboard box in and the air mattress in in a cardboard box. Note: There are pads in the case of the 15-cm thickness mattress. Also put them in the cardboard box when packing. Do not place heavy objects on the cardboard box.
  • Page 26 Max current Main fuse 1 A / 250 V Service life 2 years (based on Paramount Bed data), excluding consumable items.  Basic performance Function for supporting the user’s body weight.  Frequently used functions Hardness adjustment and alternate pressure function.
  • Page 27 Cococia Fuka must ensure that Cococia Fuka is always used in this environment. Due to its emission characteristics, Cococia Fuka is suitable for use in hospitals. If used in a home environment, this device may not offer sufficient protection against radio frequency communication services.It may be necessary for the user to take relaxation measures such as...
  • Page 28 The pump for Cococia Fūka is intended for use in the electromagnetic environments (for home and professional healthcare) specified below. The customer or the user of the pump for Cococia Fūka must use it in such an environment. * During DIP interference, the pump for Cococia Fūka will outage these normal. The cells connected with pump still have air inside which won’t affect the use and function of the system.
  • Page 29 Recommended separation distance between portable and mobile RF communications equipment and the pump for Cococia Fuka The pump for Cococia Fūka is intended for use in an electromagnetic environment (for home and professional healthcare) in which radiated RF disturbances are controlled. The customer or the user of the pump for Cococia Fūka can help prevent electromagnetic interference by maintaining a minimum...
  • Page 30 The pump for Cococia Fūka is intended to be used in an electromagnetic environment where RF radio communication devices are used. The electromagnetic interference between the RF communication device and the pump for Cococia Fūka can be prevented by maintaining the minimum distance between the two as recommended below in accordance with the maximum output and frequency of the communication device.
  • Page 31 Bed side-rails, swing arm bed grip/side grip For suitable products to be used in a combination, check the instruction manuals for beds, bed side-rails, and swing arm bed grips/side grips and Paramount Bed's catalogs and select suitable products that meet the following suitability standards.
  • Page 32: Product Symbol Definitions

    PRODUCT SYMBOL DEFINITIONS Symbol Description Symbol Description  Alternate pressure function CE mark selection Lit up: ON Indoor use Run/Standby button Lit up: ON Electrical waste Lit out: Standby Status display lamp Lit up: Please wait Refer to the instruction Flashing: Please request manual for repair work Type BF (Turns off automatically in Protection against electronic  120 minutes) shock. Lit up: ON Max inflate mode selection Alternating current (Turns off automatically in 60 minutes) Lit up: ON Protection code Be careful not to get a body part caught in the gaps Class II Equipment between the bed-side rail, (Double Insulated) swing arm bed grip, or bed and the air mattress.
  • Page 33: After-Sales Service

    Details of the fault or problem (please provide as much detail as possible) Your name, organization, address, and phone number 2. Minimum stock availability period Paramount Bed keeps replacement parts (necessary to maintain stretcher performance) in stock for at least 6 years after the discontinuation of production. 3. Questions regarding after-sales service Contact us (see "Contact Information"...
  • Page 34 MEMO...
  • Page 35 MEMO...
  • Page 36 [Original  Instruction   Manual]  AL300429 V1.01 Made by: Torindo  C o.,  Ltd.  2020-11...

This manual is also suitable for:

7f11336300a2

Table of Contents