Page 1
Turbidity systems Accurate, reliable turbidity measurement Introduction An ABB Turbidity system comprises a 4690 / 4695 analyzer and a 7998 turbidity sensor. The analyzer provides the operator interface and communications to other devices. The signal from the turbidity sensor is converted by the...
Page 2
We are an established world force in the design and manufacture of measurement products for industrial process control, flow measurement, gas and liquid analysis and environmental applications. As a part of ABB, a world leader in process automation technology, we offer customers application expertise, service and support worldwide.
Note. The following only applies to European customers. ABB is committed to ensuring that the risk of any environmental damage or pollution caused by any of its products is minimized as far as possible. The European Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/EC) that came into force on August 13 2005 aims to reduce the waste arising from electrical and electronic equipment;...
20 m (65.6 ft) A – Maximum Distance between Analyzer and Sensor 55 °C (131 °F) Max. –20 °C (–4 °F) Min. B – Within Temperature Limits IP66 C – Within Environmental Limits Fig. 3.1 Siting Requirements – Analyzer IM/4690–EN Rev. A...
Mark fixing centers (see Fig. 3.2). Position 'U' bolts on pipe. Drill suitable holes. Position plates over 'U' bolts. Secure analyzer to wall using suitable fixings. Secure plates. Secure analyzer to mounting plate. Fig. 3.3 Wall-mounting Fig. 3.5 Pipe-mounting IM/4690–EN Rev. A...
Secure the analyzer by tightening the panel clamp retaining screws. Caution. The clamp must fit flat on the analyzer casing. If the clamp is bowed, the securing screw is overtight and sealing problems may occur. Fig. 3.7 Panel-mounting IM/4690–EN Rev. A...
(12 mm [0.5 in] internal diameter) Fig. 3.8 Sensor Dimensions (With Optional Wiper Unit) Note. Allow a further 30 mm (1.2 in) (approximately) clearance above the wiper unit for the bend in the wiper unit cable. IM/4690–EN Rev. A...
An optional debubbler (part number 7997 500) is available to eliminate gas bubbles – refer to Section 3.4, page 10. Fig. 3.10 7998 Series – Sample Outlet Connector Location IM/4690–EN Rev. A...
1. To prevent degassing of the sample and very erratic readings, do not exceed this measurement. 2. This is the minimum installation distance that ensures adequate flowrate through the sensor. Increase this distance as necessary if using long or small-bore tubing. IM/4690–EN Rev. A...
AC applications or diodes for DC applications. These components can be connected either across the load or directly across the relay contacts. On 4690 / 4695 analyzers, fit the RFI components to the relay terminal block together with the supply and load wires – see Fig. 4.3.
4690 Se rie s Turbidity systems 4 Electrical Connections 4.3 Connections Mounting Terminal Number Wall Panel +12 V +12 V Cleaner Cleaner Switched emitter Cleaner / Receiver Signal input initiate pulse detect signal common supply supply Link Table 4.1 Analyzer Input Connections 4.3.1 Wall- / Pipe-mount Analyzer Connections...
Fig. 4.5 Panel-mount Analyzer Connections Note. 1. A second retransmission output is available if the RS485 serial communications facility is not used. is set to YES (see page 28), Relay 2 becomes 'Failed Wiper Alarm' relay. 2. If Test Cleaner IM/4690–EN Rev. A...
Remove front panel. Set voltage selector switch as required. 230 V AC 110 V AC Fig. 4.6 Selecting Mains Voltage – Wall- / Pipe-mount Analyzer 230 V AC 110 V AC Fig. 4.7 Selecting Mains Voltage – Panel-mount Analyzer IM/4690–EN Rev. A...
4 Electrical Connections 4.5 Turbidity Sensor Connections Blue Yellow Link Screen Emitter Blue Green Blue Green Receiver 4690 /4695 Analyzer Screen Screen Blue White Wiper Unit Screen Black (Optional) Green Yellow Fig. 4.8 Turbidity Sensor Junction Box Connections IM/4690–EN Rev. A...
New value is stored automatically Fig. 5.1 Controls and Display C – Adjusting and Storing a Parameter Value Parameter X Select New value is stored automatically D – Selecting and Storing a Parameter Choice Fig. 5.2 Membrane Key Functions IM/4690–EN Rev. A...
Manual Clean Yes is displayed for one minute. The Press the automatic clean sequence is then reset and Manual Clean No displayed. - - - - - Advance to page – see Section 7, page 20. Calibration SENSOR CAL. IM/4690–EN Rev. A...
Turbidity systems 7 Calibration 7 Calibration A key feature of the 4690 Series Turbidity system is the dry 7.1 Care and Maintenance of Secondary Standards secondary calibration standard. This is designed to simplify ABB's secondary standards may be used repeatedly but must routine calibration verification and minimizes the need for be monitored for deterioration.
(see Figs 7.2 or 7.3) facing the optical receiver, Locating ensuring the locating lug engages correctly – see Fig. 7.4. Optical Emitter Optical Receiver Fig. 7.4 Inserting Dry Standard 7. Note the reading on the display. IM/4690–EN Rev. A...
6. Insert the dry calibration standard with the zero NTU indication (see Figs 7.5 or 7.6, page 23) facing the optical receiver, ensuring the locating lug engages correctly – see Fig. 7.7. Note. Do not touch the light reflecting parts of the standard. IM/4690–EN Rev. A...
Page 25
NTU value Indication Indication Locating Optical Emitter Optical Receiver Fig. 7.5 Dry Calibration Standard for Low Range Sensors Fig. 7.7 Inserting Dry Standard NTU value Indication Zero NTU Indication Fig. 7.6 Dry Calibration Standard for High Range Sensors IM/4690–EN Rev. A...
Page 26
If Cal Fail is displayed, remove the dry standard and ensure the emitter and receiver lenses are clean and dry. Refit the dry standard, ensuring it is fitted correctly (see step 6, page 22) and repeat the procedure. Note. Do not touch the light reflecting parts of the standard. Continued on next page. IM/4690–EN Rev. A...
Page 27
Repeat the calibration procedure ensuring that an in-date or fresh formazine solution is used and that the parameter is set to the NTU value of the Fill Span Solution formazine span solution. Advance to pages – see Section 8.1, page 26. Access to Configuration IM/4690–EN Rev. A...
Language Using the keys, select the required language. Espanol Francais Deutsch English Advance to page – see Section 8.3, page 27. - - - - - Set Up Parameter SET UP PARAMETER IM/4690–EN Rev. A...
The display span also sets the correct output span. Alarm values cannot be set outside this range. Correction Factor is set to mg/l or ppm Note. Displayed only if Units Using the keys, select the required correction factor to between 0.2 and 5.0. Continued on next page. IM/4690–EN Rev. A...
Page 30
YES, set a value of 4 seconds for optimum performance. Note. If Bubble Rej. - - - - - Advance to page – see Section 8.4, page 29. Set Up Outputs SET UP OUTPUTS IM/4690–EN Rev. A...
0 to 10, 0 to 20 or 4 to 20 mA. The current output range is assigned to the selected display span – see on page 27. Display Span Continued on next page. IM/4690–EN Rev. A...
Page 32
0 0 0 0 0 Alter Calibration Security Code Set the security code to a value between 00000 and 19999. Alter Cal. Code - - - - - Advance to page. Set Up Serial Interface SERIAL INTERFACE IM/4690–EN Rev. A...
Ensure the earth on the millivolt source is connected to the earth stud. 3. Switch on the supply and allow ten minutes for the circuits to stabilize. 4. Select page and proceed as in Section 8.6.3, following. ELECTRICAL CAL IM/4690–EN Rev. A...
Set the milliammeter reading to 20 mA. Adjust RTX Span Note. The retransmission range selected in the page does not affect the reading. Set Up Outputs Return to the page – see Section 6.2.1, page 19. Operating 4 5 17 Turbidity (FNU) IM/4690–EN Rev. A...
9.1.2 Sensors With Wiper Unit The required automatic cleaning frequency of the flow chamber and optical windows of the sensors can be determined only by plant experience. It is recommended that checks are made at appropriate intervals. IM/4690–EN Rev. A...
Indicates that the contents of the nonvolatile memory has not been read correctly during power up. To rectify the problem, switch off the power, wait 10 seconds and switch on again. If the problem persists contact the Company. Table 10.1 Error Messages IM/4690–EN Rev. A...
Increasing the flowrate through the flowcell may reduce this remove short-term spikes from the readings – see page 28. For build up. Ultimately the flowcell needs to be cleaned manually. optimum operation, operate the wiper unit frequently to prevent bubbles building up on the windows. IM/4690–EN Rev. A...
Tested in accordance with IEC 61298 Parts 1-4: Edition 2.0 2008-10. Tested in accordance with MCERTS: Performance Standards and Test Procedures for Continuous Water Monitoring Equipment. Version 3.1: Environment Agency 2010. Tested in accordance with BS ISO 15839: 2003. IM/4690–EN Rev. A...
Page 40
96 mm (3.78 in.) wide x 96 mm (3.78 in.) high x 191 mm (7.52 in.) deep Weight 1.5 kg (3¼ lb) +0.8 +0.8 Panel cut-out: mm x – – +0.03 +0.03 3.62 in. x 3.62 in.) – – DS/4690–EN Rev. A IM/4690–EN Rev. A...
Cal record card kit 1 x 0219319 – Vispass bespoke 1 x STT3367 – 250 mm (10 in.) cable tie 1 x 7998140 – wiper unit 7998049 Wiper unit replacement 1 x 7998023 – wiper blade pack IM/4690–EN Rev. A...
1 x 46803002 MODBUS 7998043 7998 Turbidity EPROM (PROFIBUS) 1 x 46803003 PROFIBUS 12.6 De-bubbler Item Description Part Number O-ring large (3 off) 0211 322 O-ring small (2 off) 0211 138 Quick-fit connector (2 off) 7997 511 IM/4690–EN Rev. A...
Page 43
Operating Page Calibration Page Security Page (page 19) (page 24) (page 26) Start 45. 1 7 3. 3 74 15. 0 0 0. 0 Dry Standard Operating Parameters Formazine Electrical Calibration Page Set Up Serial Interface Page Set Up Outputs Page Set Up Parameters Page Language Page (page 33)
Page 44
Oldends Lane notice. With regard to purchase orders, the agreed — Drive Systems Stonehouse ABB Inc. particulars shall prevail. ABB does not accept any — Force Measurement Gloucestershire GL10 3TA responsibility whatsoever for potential errors or Tel: +1 215 674 6000 —...