hit counter script
Haag-Streit IM 900 Instructions For Use Manual

Haag-Streit IM 900 Instructions For Use Manual

Imaging module, accessory for the bq 900 slit lamp
Hide thumbs Also See for IM 900:
Table of Contents

Advertisement

SVENSKA
NEDERLANDS
PORTUGUÊS
ESPAÑOL
ITALIANO
FRANÇAIS
DEUTSCH
ENGLISH
INSTRUCTIONS FOR USE
Imaging Module
IM 900
Accessory for the BQ 900 Slit Lamp
5. edition / 2018 – 04
1
© HAAG‑STREIT AG, 3098 Koeniz, Switzerland – HS Doc. No. 1500.7220551.04050 – 2018 – 04
01-_IFU_IM900-7220551-04050_eng.indd 1
09.07.2018 14:10:43

Advertisement

Table of Contents
loading

Summary of Contents for Haag-Streit IM 900

  • Page 1 FRANÇAIS DEUTSCH ENGLISH INSTRUCTIONS FOR USE Imaging Module IM 900 Accessory for the BQ 900 Slit Lamp 5. edition / 2018 – 04 © HAAG‑STREIT AG, 3098 Koeniz, Switzerland – HS Doc. No. 1500.7220551.04050 – 2018 – 04 01-_IFU_IM900-7220551-04050_eng.indd 1...
  • Page 2 SVENSKA INSTRUCTIONS FOR USE Introduction Imaging Module Thank you for choosing a HAAG-STREIT device. Provided you comply carefully with the regulations in this instructions for use, we can guarantee the reliable and IM 900 unproblematic use of our product. Purpose of use Accessory for the BQ 900 Slit Lamp This device is an accessory for HAAG‑STREIT slit lamps BQ 900, which can be...
  • Page 3: Table Of Contents

    SVENSKA NEDERLANDS PORTUGUÊS ESPAÑOL ITALIANO FRANÇAIS DEUTSCH ENGLISH Contents Safety ................... 4 LED display release module RM02 ...............11 5.10 LED display camera module CM03 ..............12 Areas of application of the device ................4 5.11 Error messages (illumination head) ..............12 Ambient conditions ....................4 5.12 Error messages release module RM02..............12 Shipment and unpacking ..................4...
  • Page 4: Safety

    ENGLISH DEUTSCH FRANÇAIS ITALIANO ESPAÑOL PORTUGUÊS NEDERLANDS SVENSKA Safety Transport: Temperature from −40°C +70°C Air pressure from 500 hPa 1060 hPa DANGER! Relative humidity from Failure to comply with these instructions may result in material damage Storage: Temperature from −10°C +55°C or pose a danger to patients or users.
  • Page 5: Disinfection

    Slit lamp illumination Warranty and product liability • Haag-Streit products must be used only for the purposes and in the manner des- Plug socket on RM02 for Rotating knob on camera sym- cribed in the documents distributed with the product.
  • Page 6: System Components

    ESPAÑOL PORTUGUÊS NEDERLANDS SVENSKA System components Power supply The imaging module IM 900 is a system made up of the following main compo‑ 15. Power supply nents: 16. Mains connector (country‑dependent) A. Camera module CM03 B. Release module RM02 C. EyeSuite software D.
  • Page 7: Device Assembly / Installation

    Installation and repairs may only be performed by trained specialists. • Disassemble the breath shield. Contact your HAAG-STREIT representative for installation, repairs and • Remove the black and white cover caps (21). modification work on the system. The contact details are available at •...
  • Page 8: Cabling Of The Cm03 And The Rm02 (Diagram)

    ENGLISH DEUTSCH FRANÇAIS ITALIANO ESPAÑOL PORTUGUÊS NEDERLANDS SVENSKA Cabling of the CM03 and the RM02 (diagram) NOTE! To ensure that the system works correctly, HAAG‑STREIT recom‑ WARNING! • mends not using laptops and using a high‑quality desktop computer Only use the supplied USB 3.0 cable (3m long, HS art. no. 1022373) instead.
  • Page 9: Startup

    SVENSKA NEDERLANDS PORTUGUÊS ESPAÑOL ITALIANO FRANÇAIS DEUTSCH ENGLISH • Computer cable USB 3.0 (47), power supply cable (46) and cable headrest / LED illumination / RM02 (45) must be fed through in the braided sleeving (44). • Insert the connector plug of the power supply cable (46) in the socket (59). •...
  • Page 10: Operation

    White balance • 70% to the examiner The HAAG‑STREIT IM 900 is optimized for maximum image quality with the Switch (7) beam splitter to the top • HAAG‑STREIT slit lamp BQ 900 and BP 900. The image quality is dependent,...
  • Page 11: Conducting A White Balance

    SVENSKA NEDERLANDS PORTUGUÊS ESPAÑOL ITALIANO FRANÇAIS DEUTSCH ENGLISH 5.5.2 Conducting a white balance 5.7. LED display 8. Start the ‹EyeSuite Imaging› software illumination head 9. Activate the intensity auto mode 10. Open the ‹White balance› application 11. Start the ‹White balance› by activating the ‹Calibration› function WARNING! To achieve an optimal result during the white balance, the image must a) b) c)
  • Page 12: Led Display Camera Module Cm03

    ENGLISH DEUTSCH FRANÇAIS ITALIANO ESPAÑOL PORTUGUÊS NEDERLANDS SVENSKA 5.10 LED display camera module CM03 Green Normal operation Orange Establishing connection 5.11 Error messages (illumination head) a) b) c) Error messages Measures Incorrect supply polarisation Contact your HAAG‑STREIT representative. Illumination control not recognized Connect illumination control or replace, if necessary.
  • Page 13: Error Messages Camera Module Cm03

    SVENSKA NEDERLANDS PORTUGUÊS ESPAÑOL ITALIANO FRANÇAIS DEUTSCH ENGLISH 5.13 Error messages camera module CM03 Operational control LED (26) Error messages Measures No communication with LED illumination LI01 plus / LI02 plus Contact your HAAG‑STREIT representative. Minimum PC requirements Decommissioning Processor type Intel i5, 5th gen or higher with 4 cores.
  • Page 14: Maintenance

    We recommend having the device in‑ Statutory requirements spected by an authorized service technician annually. • The imaging module IM 900 was designed and built taking the EN 60601‑1 and Cleaning and disinfection EN 60601‑1‑2 standards into account. •...
  • Page 15: Disposal

    Electrical and electronic devices must be disposed of separately from household waste! This appliance was made available for sale after the 13th August 2005. For correct disposal, please contact your HAAG-STREIT representative. This will guarantee that no hazardous substances enter the environment and that valuable raw materials are recycled.
  • Page 16: Information And Manufacturer's Declaration Concerning Electromagnetic Compatibility (Emc)

    The operation of other lines or equipment than those listed may lead to higher emissions or may reduce the device's resistance to interfe- The Imaging Module IM 900 system fulfills the requirements on electromagnetic rence. compatibility according to EN 60601‑1‑2:2007 (IEC 3. Edition) + EN 60601‑1‑2:2015 •...
  • Page 17: Immunity (Standard Table 2)

    SVENSKA NEDERLANDS PORTUGUÊS ESPAÑOL ITALIANO FRANÇAIS DEUTSCH ENGLISH Immunity (standard table 2) The information is based on the requirements of EN 60601‑1‑2:2007 (IEC 3 edition). Guidance and manufacturer's declaration – electromagnetic immunity This product is intended for use in the electromagnetic environment specifi ed below. The customer or the user of this product should assure that it is used in such an envi- ronment.
  • Page 18: Immunity For Non-Life Support Devices (Standard Table 4)

    ENGLISH DEUTSCH FRANÇAIS ITALIANO ESPAÑOL PORTUGUÊS NEDERLANDS SVENSKA Immunity for non-life support devices (standard table 4) The information is based on the requirements of EN 60601‑1‑2:2007 (IEC 3 edition). Guidance and manufacturer's declaration – electromagnetic immunity This product is intended for use in the electromagnetic environment specifi ed below. The customer or the user of this product should assure that it is used in such an envi- ronment.
  • Page 19: Safe Distances For Non-Life Support Devices (Standard Table 6)

    SVENSKA NEDERLANDS PORTUGUÊS ESPAÑOL ITALIANO FRANÇAIS DEUTSCH ENGLISH Safe distances for non-life support devices (standard table 6) The information is based on the requirements of EN 60601‑1‑2:2007 (IEC 3 edition). Recommended safe distances between portable and mobile HF communication devices and this device. This product is designed to be operated in an electromagnetic environment in which radiated HF interference is controlled.
  • Page 20 FRANÇAIS ITALIANO ESPAÑOL PORTUGUÊS NEDERLANDS SVENSKA Should you have any further questions, please contact your HAAG‑STREIT dealer at: http://www.haag-streit.com/contact/contact-your-distributor.html HAAG-STREIT AG Gartenstadtstrasse 10 3098 Koeniz, Switzerland Phone +41 31 978 01 11 +41 31 978 02 82 eMail info@haag-streit.com Internet www.haag-streit.com...

Table of Contents